COMPARATIVE ANALYSIS OF CYCLIC FATIGUE FRACTURE RESISTANCE OF DIFFERENT TYPES OF ENDODONTIC FILES
DOI:
https://doi.org/10.51891/rease.v12i11.31892Keywords:
Endodontics, Cyclic fatigue, FractureAbstract
This study analyzed the cyclic fatigue fracture resistance of the Logic® (Easy), Sequence® (MK Life), and Wave One Gold® (Dentsply Sirona) endodontic systems. The files underwent cyclic fatigue testing in resin blocks containing simulated canals with double curvature, a 6 mm radius, and angles of 30° and 45°. The time elapsed until fracture for each instrument was recorded, and the number of cycles to fracture (NCF) was calculated. The Sequence® system exhibited the highest mean values for time and NCF, while Wave One Gold® showed longer times but greater variability among specimens. Logic® demonstrated intermediate performance and greater result consistency. However, the differences observed between the groups were not statistically significant. The evaluated systems showed similar performance regarding cyclic fatigue resistance. No statistically significant influence was observed regarding the time to fracture, NCF, or the length of the fractured fragment.